Mass-transfer simulation test and theoretical model analysis in dialysis hollow fiber membrane |
| Authors: Hu Xianghua1,2, Zhang Lizhi1, Lan Jianhua2, Luo Jiewei2 |
| Units: 1 Guangdong Medical Instrument Supervision and Test Center of State Food Drug Administration, Guangzhou,Guangdong,510663; |
| KeyWords: dialysis hollow fiber membrane;mass transfer model; diffusion coefficient |
| ClassificationCode:TQ021.4 |
| year,volume(issue):pagination: 2015, 35(1):70-74 |
|
Abstract: |
|
Hollow fiber dialyzer is widely applied in the end-stage renal disease in hemo-dialysis therapy. It plays an important role in kidney disease patients in the world. However, there is little research of solute transfer in dialyzer. This paper builds a theoretical mass-transfer model of dialysis hollow fiber membrane, based on mass-conservation theory and double-membrane theory. Three common membrane materials are used in simulation test,and data is calculated with the mass transfer model with represent solutes. With analysis of the relationship between mass-transfer performance and the structural characteristics of membrane, referential suggestions for material selection and structural design in manufacture of hollow fiber membrane are put forward. |
|
Funds: |
| 广州市科研条件建设项目(穗科信字[2013]163-5号) |
|
AuthorIntro: |
| 胡相华,女,广东人,工程硕士,医疗器械工程师,研究方向:医疗器械/化学工程,联系方式Email:oab613@sohu.com. |
|
Reference: |
|
[1] 中国医院协会血液净化中心管理分会血液透析登记组. 我国面临快速增长的终末期肾病治疗负担[J]. 中国血液净化, 2010,9(1):47~49. |
|